Customization: | Available |
---|---|
Type: | Ventilator Machine |
Used for: | Adult, Child, Infant |
Still deciding? Get samples of US$ 15000/Piece
Request Sample
|
Suppliers with verified business licenses
Audited by an independent third-party inspection agency
Ventilation modes: | IPPV, A/C, PCV, SIMV, SPONT/CPAP, PSV |
Ventilator parameter ranges |
|
Tidal volume(Vt) | 0 ~ 2000mL |
Frequency (Freq) Oxygen concentration |
1bpm ~ 100 bpm 21%-100% |
I/E | 4:1~1:8 |
PEEP | 0cmH2O ~ 40 cmH2O |
Pressure Limit | 20 cmH2O ~ 100 cmH2O |
Monitored Parameters |
|
Frequency (Freq) | 0 /min ~ 100 /min |
Tidal volume(Vt) | 0 mL ~ 2500 mL |
MV Airway pressure Dynamic lung compliance testing ETCO2 Concentration |
0 L/min ~ 99 L/min 0 cmH2O ~ 100 cmH2O 1mL/cmH2O ~ 1000mL/cmH2O 0mmHg ~ 152mmHg (0%-20%) |
Oxygen concentration | 15 % ~ 100 % |
Oscillographs display: |
|
P-T(pressure - time) | |
F-T(flow - time) | |
P-V loop (pressure - volume loop) | |
V-T(volume - time ) | |
ETCO2-T (End-tidal CO2 - time) | |
Alarms and protection |
|
The AC power failure alarm | Power failure or no connection |
Internal battery backup low voltage alarm | ≤11.3±0.3V |
No tidal volume | No tidal volume within 6s |
High Minute Volume alarm Low Minute Volume alarm |
5L/min-99L/min 1L/min-30L/min |
High Airway pressure alarm Low Airway pressure alarm High oxygen concentration alarm Low oxygen concentration alarm |
20cmH2O-100cmH2O 0cmH2O-20cmH2O 19%-100% 18%-99% |
Continuous pressure alarm | (PEEP+1.5cmH2O) over 16s |
Suffocation warning Fan error Oxygen deficit |
5-60s Show on screen Show on screen |
The maximum limited pressure | <12.5 kPa |
Working conditions |
|
Gas source | O2, AIR |
Pressure | 280kPa-600kPa |
Voltage | ~220V±22V |
Power frequency | 50Hz ± 1Hz |
Input power | 900VA (with air compressor) 250VA (without air compressor) |